2011/03/21

0840 Stop flashing detector
0850 Fill N2 cold trap full
1314 The data took yesterday can not be processed by NoiseGen because file header error. Take some 500-event files to test cdmsbats.
1322 NoiseGen worked after changing the P and Q sample number.
1323 Start to take data 07110321_1323_F0001. Trace length 0.4096ms, lookback time 100us. Trigger coincidence time 50us. -4V on Q1, 0V on Q2-4. Trigger: 64ADC on Q1. Tower temperature 60.4mK.
1742 Stop the run at 10th file and flash the detector.
1840 Stop flashing.
1943 Start run 07110321_1943_F0001. Trace length 0.4096ms, lookback time 100us. Trigger coincidence time 50us. -4V on Q1-4. Trigger: 64ADC on Q1. Ba-133 souce. Tower temperature 61.3mK.
2026 Fill N2 cold trap full.
2255 Flashing DCRC Boards. Stopped and restarted to make sure flashing was done correctly.
2328 Board temperature at 565 mk.
2345 Stopped flashing boards.