| 0852 | Stop flashing the detector. |
| 0914 | Fill N2 cold trap full |
| 1125 | Start to take data 07110319_1125_F0001. 2000 random triggers. -6V on Q1, 0V on Q2-4. Trigger, 64ADC on Q1. Ba-133 source. Tower: 63.4mK |
| | Stop the because the trigger didn't seem to happen at the right place. |
| | Resynchronize the two boards. |
| 1130 | start the run 07110319_1130_F0001. 2000 random triggers. -6V on Q1, 0V on Q2-4. Trigger, 64ADC on Q1. Ba-133 source. Tower: 63.4mK |
| 1134 | The trigger worked OK now. The random trigger was set to 2000 instead of 500. |
| 1624 | Stop the run at 07110319_1130_F0014. Flash the detector. |
| 1837 | Stop flashing |
| 1924 | Start take data 07110319_1924. 500 random triggers. -4V on Q1, 0V on Q2-4. Trigger, 40ADC on Q1. Ba-133 source. |
| 2214 | Stop the run 07110319_1924 at 9th file. I feel kind of suffocated staying in the lab with ventilation off. I think the reason why we could not see Am/Ba peaks is because there are too many pileup events. Shortening the trace length may be able to fix this. |
| 2217 | Flash the detector. |