2011/03/16
| 0025 | Open gate valve to check flanges behind it that lie outside the fridge. He leak rate rises from 6.5E-7 mbar l/s to 7.5 mbar l/s. |
|---|---|
| 0105 | Spray test all flanges behind gate valve and condenser line speedivalve. He leak rate falls to 6.5E-7 mbar l/s. No leak. |
| Close all valves and isolate leak checker in preparation for removal from vent on IGH. | |
| 0130 | Leak checker reconnected to IVC. Isolate leak checker and bellows, and leave to pump overnight to reduce helium background. |
| 0135 | Fill N2 cold trap full. |
| Pump out helium rubber tube with He-4 pump. Close vent speedivalve, and valves 2a and 5a on IGH. Switch off He-4 pump. | |
| 0150 | Start cryocleaning mixture through cold trap 2. Valve 6 open 13.7%. |
| 0240 | Flow seems to have stopped. Open valve 6 14.0% while waiting for bus. |
| 0936 | IVC: 4.6mbar. Pump out N2 exchange gas. |
| 0948 | Electrical check done. Everything's ok. |
| 0958 | Pumped Mixture into dump. |
| 1008 | Opened Gate Valve and Condenser SpeediValve. |
| 1024 | Performed through-put test. Result is 40 mbar/ minute. |
| 1025 | No leak from circulation line to IVC. |
| 1049 | Blow out of LN2 begun. |
| 1207 | G3 start to drop. Liquid N2 is out. Leak signal start to increase from 1.2e-8cc/s |
| 1211 | Shut off He gas |
| 1224 | IVC He signal 1.4E-8cc/s. |
| 1231 | IVC He BG 1.45E-8cc/s. |
| 1232 | Start to pump the bath. |
| 1237 | P2 start to drop. Liquid was all out. |
| 1245 | IVC He signal 1.44e-8cc/s. There should not be leak between the bath and IVC. |
| 1246 | Fill bath with He gas |
| 1249 | G3 started to rise. Close V2A, V5A to stop the He gas. IVC He signal: 1.43e-8cc/s. |
| 1252 | Turned of V1A; turned on He 4 Pump; closed vent SpeediValve. |
| 1304 | Evacuated vent line (turning of V2a, V5a); Evacuated 1k pot and isolated it (opened and closed the big speedi valve) |
| 1308 | Closed off IVC. |
| 1321 | Filled IVC with 0.1 mbar of He4 exchange gas. There was 2.5 mbar in the pumping line before IVC was opened. Ready to transfer liquid Helium |
| 1340 | Begun transfer of Liquid Helium. |
| 1603 | LS370 overranged at 2.5kOhm, 4.2K on X12004. |
| 1615 | Start to pump out exchange gas. MC X12004: 4.2 K. It's a bit too late to pump out the exchange gas. |
| 1627 | Tower temperature: 8K |
| 1702 | Helium transfer tube removed. Meter readings and helium level recorded. |
| 1731 | Electrical check- good. |
| Cooldown checkout sheet (rev C SQUET) | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Detector: | G101 DIB1 | BH: | Stripline: | 19 | |||||||
| SC: | ZB1-1-9 | 300 K | LN | LHe | |||||||
| Face: | 5 | Date | 2011-03-12 | Date | 2011-03-13 | Date | 2011-03-16 | ||||
| SQUET: | ZC23 | Who | HC & JZ | Who | JZ & HC | Who | RR | ||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | (Gnd lead) | |||||||||
| Nominal: | Nominal: | Nominal: | |||||||||
| < 0.1 nS | < 0.1 nS | < 0.1 nS | |||||||||
| Phonon | |||||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.4 | OK | 1 - 2 k | 1.6 | OK | 80 - 140 | 101.6 | OK |
| SQUID bias B/Ret | 19 | 34 | 5 | OK | 2 | OK | 80 - 140 | 89.2 | OK | ||
| SQUID bias C/Ret | 8 | 24 | 4.6 | OK | 1.7 | OK | 80 - 140 | 99.2 | OK | ||
| SQUID bias D/Ret | 23 | 39 | 4.3 | OK | 1.6 | OK | 80 - 140 | 89.3 | OK | ||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 19 | OK | 9 - 13 k | 7.3 | OK | 1.0 - 1.3 k | 1.28 | OK |
| Feedback B/Ret | 2 | 18 | 21 | OK | 8.2 | OK | 1.0 - 1.3 k | 1.28 | OK | ||
| Feedback C/Ret | 41 | 25 | 19 | OK | 7.6 | OK | 1.0 - 1.3 k | 1.27 | OK | ||
| Feedback D/Ret | 22 | 6 | 18 | OK | 7 | OK | 1.0 - 1.3 k | 1.27 | OK | ||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 94,87,85,86,86 | OK | 30 - 40 | 50,45,43,44,44 | OK | 20 - 30 | 32.3,28.3,26.5,27.5,27.4 | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 92,92,100,92 | OK | 44,44,47,44 | OK | 20 - 30 | 26.9,26.5,29.7,26.6 | OK | ||
| QET C/Ret | 7 | 5, 9, 10, 28 | 87,85,74,84 | OK | 45,44,33,44 | OK | 20 - 30 | 28.4,27.6,17,26.7 | OK | ||
| QET D/Ret | 38 | 36, 40, 43, 44 | 85,84,84,84 | OK | 44,44,43,44 | OK | 20 - 30 | 27.9,27.2,26.7,27 | OK | ||
| Charge | |||||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.068 | OK | 1.0 - 1.1 k | 1.043 | OK | 1.0 - 1.1 k | 1.23 | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | 26 (PA), open (PB, PC, PD) | OK | < 1 nS | >50M, OL | OK | <1 nS | OK | OK | |
| Qo bias/QET Ret | 11 | 26 (PA), open (PB, PC, PD) | OK | >50M, OL | OK | <1 nS | OK | OK | |||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 23 | OK | 6 - 8 nS | 20 | OK | <1 nS | 54.46 MΩ, 54.56 MΩ | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 23 | OK | 20 | OK | <1 nS | 52.5 MΩ, 52.3 MΩ | OK | ||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK | <0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qi source | 31 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qo source | 27 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.721 | OK | 0.9 - 1.0 V | 1.04 | OK | — | 1.9 | OK |
| LED 1 | 42 | chassis | 1.0 - 1.1 V | 1.137 | – | 1.2 - 1.5 V | 1.418 | – | — | OL | – |
| LED 2 | 26 | chassis | open | – | open | – | — | OL | – | ||
DIB 2
| Cooldown checkout sheet (rev C SQUET) | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Detector: | G101 DIB2 | BH: | Stripline: | 20 | |||||||
| SC: | ZB1-1-6 | 300 K | LN | LHe | |||||||
| Face: | 2 | Date | 2011-03-12 | Date | 2011-03-13 | Date | 2011-03-16 | ||||
| SQUET: | ZC70 | Who | HC & JZ | Who | JZ & HC | Who | RR | ||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | (Gnd lead) | |||||||||
| Nominal: | Nominal: | Nominal: | |||||||||
| < 0.1 nS | < 0.1 nS | < 0.1 nS | |||||||||
| Phonon | |||||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.9 | OK | 1 - 2 k | 1.9 | OK | 80 - 140 | 95 | OK |
| SQUID bias B/Ret | 19 | 34 | 4.4 | OK | 1.6 | OK | 80 - 140 | 99.8 | OK | ||
| SQUID bias C/Ret | 8 | 24 | 4.7 | OK | 1.8 | OK | 80 - 140 | 91.4 | OK | ||
| SQUID bias D/Ret | 23 | 39 | 5.3 | OK | 2.1 | OK | 80 - 140 | 91 | OK | ||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 29 | OK | 9 - 13 k | 15 | OK | 1.0 - 1.3 k | 1.26 | OK |
| Feedback B/Ret | 2 | 18 | 26 | OK | 9 | OK | 1.0 - 1.3 k | 1.26 | OK | ||
| Feedback C/Ret | 41 | 25 | 29 | OK | 14 | OK | 1.0 - 1.3 k | 1.27 | OK | ||
| Feedback D/Ret | 22 | 6 | 35 | OK | 17 | OK | 1.0 - 1.3 k | 1.25 | OK | ||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 89,82,80,80,80 | OK | 30 - 40 | 50,44,42,43,42 | OK | 20 - 30 | 25.5,21.1,19.1, 20.1, 19.6 | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 83,82,90,83 | OK | 37,36,41,36 | OK | 20 - 30 | 20.6,20.1,23.3,20.3 | OK | ||
| QET C/Ret | 7 | 5, 9, 10, 28 | 79,77,69,76 | OK | 41,40,32,40 | OK | 20 - 30 | 20.9,19.9,12.4,19.3 | OK | ||
| QET D/Ret | 38 | 36, 40, 43, 44 | 81,80,80,80 | OK | 43,43,42,43 | OK | 20 - 30 | 21.6,21.1,20.7,20.9 | OK | ||
| Charge | |||||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.061 | OK | 1.0 - 1.1 k | 1.04 | OK | 1.0 - 1.1 k | 1.23 | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | open | OK | < 1 nS | open | OK | <1 nS | open | OK | |
| Qo bias/QET Ret | 11 | open | OK | open | OK | <1 nS | open | OK | |||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 23,23 | OK | 6 - 8 nS | 20,20 | OK | <1 nS | ok | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 23,23 | OK | 20,20 | OK | <1 nS | ok | OK | ||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK | <0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qi source | 31 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qo source | 27 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.706 | OK | 0.9 - 1.0 V | 1.082 | OK | — | 1.6 | OK |
| LED 1 | 42 | chassis | 1.0 - 1.1 V | 1.129 | – | 1.2 - 1.5 V | 1.412 | – | — | OL | – |
| LED 2 | 26 | chassis | 1.121 | – | 1.405 | – | — | OL | – | ||
| 1829 | Close IVC, start condensing mixture. |
|---|---|
| 1942 | G1 = 81.4 mbar, G2 = 83 mbar and both plateau. Stop condensing. |
| 1945 | Start to circulate mixture. Tower T = 2.3K. |
First Pulses
| 2130 | Flash detector overnight: DIB1 LED1, 0.6 mA, 100 μs ON, 2500 μs OFF; DIB2 LED1&2, 0.6 mA, 100 μs ON, 2500 μs OFF. |
|---|---|
| 2245 | Fill N2 cold trap full. |