2011/03/13
| 1745 | Fill N2 cold trap full. |
|---|---|
| Mixing chamber temperature 78.6 K, falling gradually. Probably alright to pump out N2 exchange gas, but have other jobs to do so no harm in delaying it. | |
| 1830 | Tower temperature 78.3 K. Pump out N2 exchange gas. |
| 1920 | Conduct electrical check at 77 K. |
| Detector: | G101 DIB1 | BH: | Stripline: | 19 | ||||
| SC: | ZB1-1-9 | 300 K | LN | |||||
|---|---|---|---|---|---|---|---|---|
| Face: | 5 | Date | 2011-03-12 | Date | 2011-03-13 | |||
| SQUET: | ZC23 | Who | HC & JZ | Who | JZ & HC | |||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | |||||||
| Nominal: | Nominal: | |||||||
| < 0.1 nS | < 0.1 nS | |||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.4 | OK | 1 - 2 k | 1.6 | OK |
| SQUID bias B/Ret | 19 | 34 | 5 | OK | 2 | OK | ||
| SQUID bias C/Ret | 8 | 24 | 4.6 | OK | 1.7 | OK | ||
| SQUID bias D/Ret | 23 | 39 | 4.3 | OK | 1.6 | OK | ||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 19 | OK | 9 - 13 k | 7.3 | OK |
| Feedback B/Ret | 2 | 18 | 21 | OK | 8.2 | OK | ||
| Feedback C/Ret | 41 | 25 | 19 | OK | 7.6 | OK | ||
| Feedback D/Ret | 22 | 6 | 18 | OK | 7 | OK | ||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 94,87,85,86,86 | OK | 30 - 40 | 50,45,43,44,44 | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 92,92,100,92 | OK | 44,44,47,44 | OK | ||
| QET C/Ret | 7 | 5, 9, 10, 28 | 87,85,74,84 | OK | 45,44,33,44 | OK | ||
| QET D/Ret | 38 | 36, 40, 43, 44 | 85,84,84,84 | OK | 44,44,43,44 | OK | ||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.068 | OK | 1.0 - 1.1 k | 1.043 | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | 26 (PA), open (PB, PC, PD) | OK | < 1 nS | >50M, OL | OK | |
| Qo bias/QET Ret | 11 | 26 (PA), open (PB, PC, PD) | OK | >50M, OL | OK | |||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 23 | OK | 6 - 8 nS | 20 | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 23 | OK | 20 | OK | ||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | – | OK | ||
| Qi source | 31 | chassis | – | OK | – | OK | ||
| Qo source | 27 | chassis | – | OK | – | OK | ||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.721 | OK | 0.9 - 1.0 V | 1.04 | OK |
| LED 1 | 42 | chassis | 1.0 - 1.1 V | 1.137 | – | 1.2 - 1.5 V | 1.418 | – |
| LED 2 | 26 | chassis | open | – | open | – | ||
| Detector: | G101 DIB2 | BH: | Stripline: | 20 | ||||
| SC: | ZB1-1-6 | 300 K | LN | |||||
|---|---|---|---|---|---|---|---|---|
| Face: | 2 | Date | 2011-03-12 | Date | 2011-03-13 | |||
| SQUET: | ZC70 | Who | HC & JZ | Who | JZ & HC | |||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | |||||||
| Nominal: | Nominal: | |||||||
| < 0.1 nS | < 0.1 nS | |||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.9 | OK | 1 - 2 k | 1.9 | OK |
| SQUID bias B/Ret | 19 | 34 | 4.4 | OK | 1.6 | OK | ||
| SQUID bias C/Ret | 8 | 24 | 4.7 | OK | 1.8 | OK | ||
| SQUID bias D/Ret | 23 | 39 | 5.3 | OK | 2.1 | OK | ||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 29 | OK | 9 - 13 k | 15 | OK |
| Feedback B/Ret | 2 | 18 | 26 | OK | 9 | OK | ||
| Feedback C/Ret | 41 | 25 | 29 | OK | 14 | OK | ||
| Feedback D/Ret | 22 | 6 | 35 | OK | 17 | OK | ||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 89,82,80,80,80 | OK | 30 - 40 | 50,44,42,43,42 | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 83,82,90,83 | OK | 37,36,41,36 | OK | ||
| QET C/Ret | 7 | 5, 9, 10, 28 | 79,77,69,76 | OK | 41,40,32,40 | OK | ||
| QET D/Ret | 38 | 36, 40, 43, 44 | 81,80,80,80 | OK | 43,43,42,43 | OK | ||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.061 | OK | 1.0 - 1.1 k | 1.04 | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | open | OK | < 1 nS | open | OK | |
| Qo bias/QET Ret | 11 | open | OK | open | OK | |||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 23,23 | OK | 6 - 8 nS | 20,20 | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 23,23 | OK | 20,20 | OK | ||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | – | OK | ||
| Qi source | 31 | chassis | – | OK | – | OK | ||
| Qo source | 27 | chassis | – | OK | – | OK | ||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.706 | OK | 0.9 - 1.0 V | 1.082 | OK |
| LED 1 | 42 | chassis | 1.0 - 1.1 V | 1.129 | – | 1.2 - 1.5 V | 1.412 | – |
| LED 2 | 26 | chassis | 1.121 | – | 1.405 | – | ||
| 1835 | IVC: 40mbar, MC X12004: 78.3K. Start to pump out N2 exchange gas. |
|---|---|
| 1908 | Finished electrical check: OK |
| 1911 | IVC: 1.9e-3 mbar. Isolate pumping station, He BG ~3e-9 cc/s. Connect pumping station and keep pumping. |
| 1955 | Pump mixture into dump to conduct throughput test. |
| 2040 | Open gate valve, condenser speedivalve. |
| 2043 | Conduct throughput test: 15 mbar/min. No leak from circulation line. Pump mixture into dump to try again. |
| 2047 | Pump mixture into dump. |
| 2107 | Conduct another throughput test: 12 mbar/min. Pump mixture into dump again. |
| P1 easily goes down to 0.007 mbar. There might be a partial blockage in the circulation line. | |
| 2124 | Close gate and condenser line speedivalves. |
| 2140 | Start cryocleaning mixture again. Valve 6 open 25.7%: quite high, but necessary to get a good flow rate. Why? |
| 2144 | Open valve 6 to 25.8%. |
| 2158 | Open valve 6 to 25.9%. Can see frost on N2 cold trap outlet: maybe too fast. |
| 2303 | Pump mixture into dump. |
| 2345 | Conduct throughput test: 19 mbar/min. There is probably a blockage. Might need to warm up fridge. Pump mixture into dump. |