2011/03/10

1400 Mill second Schuba extender plug.
2000 Take apart tower to extract sources from source holder for audit.
Try to photograph LEDs while applying a current of 1 μA. Cannot see LEDs glowing, probably due to infrared filter on camera lens.
Perform detector level continuity checks of LED channels on other detectors lying in purge cabinet.
Table 1: Detector level continuity checks of LED channels
Signal G8 G9 S11C G51 DIB1 G51 DIB2
LED1 Rforward 698 kΩ 749 kΩ 752 kΩ 788 kΩ 793 kΩ
LED2 Rforward 715 kΩ 745 kΩ 748 kΩ 750 kΩ 792 kΩ
LED1 Vforward 1.062 V 1.058 V 1.057 V 1.063 V 1.062 V
LED2 Vforward 1.067 V 1.060 V 1.058 V 1.062 V 1.062 V
LED1 Rreverse 0.2 nS 0.1 nS 0.2 nS 0.2 nS <0.1 nS
LED2 Rreverse 0.4 nS 0.1 nS 0.7 nS 3.0 nS <0.1 nS
2100 Place tower in N2 purge cabinet.