2011/03/08

Figure 1: Orientation of source holder and sources above crystal.
mountsourceholder.jpg
0100 Place tower (G101 detector and source holder attached) in N2 purge cabinet.

Modifications to Schuba Extender Plugs

1600 Use mill to shave 0.14” of copper off Schuba extender plug.
M2 screws connecting Schuba extender plug pins to copper piece are too long. Replaced with M2 x 6 mm flat head screws.
Due to reducing the thickness of the Schuba extender plug, the 0-80 x 1” screws are too long to fasten the side coax and plug to the tower. Replaced with shorter brass ones lying around the lab.
Figure 2: Original and modified Schuba extender plugs side-by-side.
compareschubas.jpg
Figure 3: The modified extender now sits flush with the tower.
schubatowerfit.jpg
Figure 4: Side coax now fits nicely.
schubasidecoaxfitall.jpg
Figure 5: Close-up view of side coax connected to DIB1 and Schuba extender plug.
schubasidecoaxfittop.jpgschubasidecoaxfitbottom.jpg

Tower Level Continuity Checks

2000 Tower level continuity check performed with Fluke 177 DMM.
DMM can measure maximum resistance of 50 MΩ (ie. 20 nS). Fluke 87 DMM will arrive tomorrow.
LED2 open at detector level. Not sure what happened.
Table 1: DIB1 tower level continuity check with Fluke 177 DMM.
Signal Measured
QET A 126.3 Ω
LED1 Rforward 769 kΩ
LED2 Rforward open
LED1 Vforward 1.106 V
LED2 Vforward open
LED1 Rreverse <20 nS
LED2 Rreverse <20 nS
Qo Rbias 26 nS
Qo Ccoupling <20 nS
Qo Rfeedback 26 nS
Qi Rbias 26 nS
Qi coupling <20 nS
Qi Rfeedback 26 nS