start › cdms › k100 › run_diary › 20110226run011

2011/02/24

1500 Brought box containing puck and crystal to laboratory.
Opened box. Was not expecting that…
Pressure of N2 within puck = 8 psig.
openbox.jpg

2011/02/25

1440 Remove puck from elastic bands.
1500 Puck cleaned and taken to clean room.

2011/02/26

Procedure for Removing 100 mm Crystal from Shipping Container

1255 Crystal removed from puck as per Gary's instructions.
Figure 1: Accelerometer removed from top of puck.
removeaccelerometer.jpg
Figure 2: Pressure released with fitting. Apologies for the poor quality.
releasepressure.jpgfittings.jpg
Figure 3: Unscrewed six socket head 8-32 x ¾“ bolts with 9/64” hex key.
unscrewlid.jpg
Figure 4: Do not be alarmed! This photo was taken after one of the four slot head
steel 4-40 screws holding the turntable to the floor of the puck was unscrewed.
unscrewfloor.jpg
Figure 5: Crystal lifted out of puck and turned over. Removed four slot head steel 4-40 screws securing turntable to aluminium adapter plate.
turntable.jpgunscrewturntable.jpg
Figure 6: Lifted off turntable to expose aluminium plate. Removed six Phillips pan head M3 brass screws securing the Al adapter plate to the housing's copper lid.
aladapter.jpgunscrewaladapter.jpg
Figure 7: Crystal ready to be mounted. Placed in purge cabinet.
upsidedowncrystal.jpg

Detector Checkout

1330 Detector level continuity check performed. See detector e-log entry for previous values.
Table 1: Detector level continuity checks
DIB 1 DIB 2
Signal Measured Measured
QET A 36.6 Ω N/A
LED1 R forward 774 kΩ 795 kΩ
LED1 V forward 1.062 V 1.062 V
LED1 R backward <1 nS <1 nS
LED2 R forward 787 kΩ 790 kΩ
LED2 V forward 1.063 V 1.062 V
LED2 R backward <1 nS <1 nS
Qo(+ve)-Qi(-ve) 243 Ω 264 Ω
Qi(+ve)-Qo(-ve) 210 Ω 236 Ω

Data from Accelerometer

1500 Plug accelerometer into computer. Stop recording and download data.
Data available in csv format from here.
Table 2: Vertical red lines in Figures 8, 9 & 10 correspond to the following events from FedEx's website in chronological order (tracking no. 057573930010329):
Date Time [PST] Activity Location
Feb 18 2011 14:07 Picked up San Jose, CA
Feb 18 2011 18:40 Arrived at FedEx location San Jose, CA
Feb 18 2011 23:00 Left FedEx origin facility San Jose, CA
Feb 19 2011 04:57 Arrived at FedEx location Sacramento, CA
Feb 19 2011 09:03 Departed FedEx location Sacramento, CA
Feb 21 2011 20:27 Arrived at FedEx location North Salt Lake, UT
Feb 22 2011 04:45 Departed FedEx location North Salt Lake, UT
Feb 23 2011 13:31 Arrived at FedEx location St. Paul, MN
Feb 24 2011 02:40 At local FedEx facility St. Paul, MN
Feb 24 2011 04:03 On FedEx vehicle for delivery St. Paul, MN
Feb 24 2011 07:05 Delivered Minneapolis, MN
Figure 8: Pressure exerted on puck during its journey across the country as measured by the accelerometer.
Figure 9: Temperature of puck over the course of its voyage across the country as measured by the accelerometer.
Figure 10: G-force sustained by puck during its traumatic odyssey as measured by the accelerometer. Note the different x-axis range to Figures 8 & 9 above. The lines between the points have no meaning as the accelerometer recorded the acceleration when its trigger threshold was satisfied. They are present here to help your eyes.