| 1200 | Stop flashing detectors. |
| | Fill N2 cold trap Full. |
| 1312 | The temperature could not go lower than 70mK when the still heater is off. G1 and G2 ~50mbar and P1~0.05mbar. The still seemed too cold. Add 6mW to the still heater. Tower temperature starts to oscillate. |
| 1515 | Tower temperature 88 mK. Turn off Still. |
| 1555 | Try to optimise Still: 4 mW. Temperature starts to drop: reaches 78 mK. Pressure P1 = 0.09 mbar. |
| 1405 | Turn power of Still Heater down to 3 mW. Temperature starts to rise towards 79 mK, pressure P1 = 0.08 mbar. |
| | Still temperature high at 1.2 K. Turn off Still heater and open valve 14 to allow more mixture out of dump into Still. P1 rises to 1.5 mbar. |
| 1415 | Close valve 14. Pressure P1 drops to 0.04 mbar. Tower temperature dropping; 78 mK at the moment. |
| 2020 | Tower temperature 71.4 mK. Plug in DCRC boards; temperature rises to 82 mK before returning to original value. |
| 2300 | Changed the firmwares two boards to the one sent by Sten recently with checksum FC33. And the charge baseline oscillation noise of DCRC33 was reduced. |