| 1235 | Fill N2 cold trap full. |
| 1530 | Stop cryocleaning mixture. Pump mixture into dump. |
| 1535 | Start 1 K pot. |
| 1550 | Start condensing mixture. Adjust needle valve until 1 K pot temperature is 1.773 K. |
| 1745 | Tower temperature 3.2 K. Stop condensing mixture; start circulation. |
| 1955 | Turn Still on: 6 mW. |
| 2010 | Tower temperature 82.7 mK. Seems to be stable. Wait for a while. |
| 2030 | Flash G9: LED1; current 500 μA; flash duration 100 μs; flash-off time 2.56 ms. Tower temperature peaks at 142 mK. |
| 2040 | Flash G8: LED1; current 600 μA; flash duration 100 μs; flash-off time 2.56 ms. Tower temperature peaks at 246 mK. Tower temperature did not increase by very much when current was at a 500 μA. Looking at the traces, both detectors are being flashed. Leave to flash overnight. |
| 2145 | Liquid helium level jumped from 17.2% to 17.6%. There seem to be some fluctuations, so these values can't be trusted. Seem to be losing 2% every hour, so might run out by 06:00 tomorrow. |
| | Start LHe transfer. |
| 2230 | Stop LHe transfer. Helium liquid level 96.1%. |
| 2250 | Switch off Still. |
| | Fill N2 cold trap full. |