2011/01/31

1000 Fill N2 cold trap full.
Tower temperature stable at 155 mK. Pressure in G2 high at 180 mbar.
1400 Tower temperature seems to rise. Could this be due to the DCRC boards?
1620 Tower temperature rising. Helium level 5.8%. Perform LHe transfer.
1730 LHe transfer complete, helium level 96.1%. Tower temperature dropping.
1800 Tower temperature reaches 165 mK. Things seem to be back to normal.
2000 Tower temperature increasing inexplicably. Pressure in G2 is 96 mbar. Turn on Still (5.0 mW) to increase circulation and increase needle valve to 1 K pot. Increase causes 1 K pot temperature to rise from 1.6 K to 1.7 K. Why?
2015 Too early to turn on Still: tower temperature increases. Turn Still off. Switch to nitrogen cold trap 2 as cold trap 1 might be dirty.
2130 Tower temperature hovering around 220-230 mK. Pressure in G2 119 mbar. Temperature of 1 K pot is 1.7 K, and exhaust line shows a relatively high pressure of 50 mbar. Cold trap 2 seems to be helping somewhat. Perhaps cold trap 1 needs a more thorough clean.
2330 Base temperature still rising. Noticed helium trap was loose, so tightened it. This led to a reduction in temperature. I have no idea why.