2011/01/25
| 1400 | Stop pumping IVC at pressure of 7.4E-6 mbar |
|---|---|
| Background leak rate of 1E-8 mbar l/s | |
| Pump dewar and flush with He gas: leak rate of 1.2E-8 mbar l/s | |
| 1500 | Evacuate 1 K pot |
| 1510 | Flood IVC with 4.6 mbar of N2 exchange gas (35 mbar prior to isolating IVC) |
| 1645 | Begin LN2 transfer |
| 1900 | Electrical check at 220 K |
Electrical Checks
| Cooldown checkout sheet (rev C SQUET) | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Detector: | G9 | BH: | Stripline: | 19 | |||||||
| SC: | ZB1-4-6 | 220 K | LN | LHe | |||||||
| Face: | 2 | Date | 2011-01-25 | Date | 2011-01-26 | Date | |||||
| SQUET: | ZC23 | Who | HC | Who | HC | Who | |||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | (Gnd lead) | |||||||||
| Nominal: | Nominal: | Nominal: | |||||||||
| < 0.1 nS | < 0.1 nS | < 0.1 nS | |||||||||
| Phonon | |||||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 2.7 | OK | 1 - 2 k | 1.6 | OK | 80 - 140 | ||
| SQUID bias B/Ret | 19 | 34 | 3.2 | OK | 2 | OK | 80 - 140 | ||||
| SQUID bias C/Ret | 8 | 24 | 2.9 | OK | 1.7 | OK | 80 - 140 | ||||
| SQUID bias D/Ret | 23 | 39 | 2.7 | OK | 1.6 | OK | 80 - 140 | ||||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 12 | OK | 9 - 13 k | 7.3 | OK | 1.0 - 1.3 k | ||
| Feedback B/Ret | 2 | 18 | 13 | OK | 8.2 | OK | 1.0 - 1.3 k | ||||
| Feedback C/Ret | 41 | 25 | 12 | OK | 7.6 | OK | 1.0 - 1.3 k | ||||
| Feedback D/Ret | 22 | 6 | 12 | OK | 7.1 | OK | 1.0 - 1.3 k | ||||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 54,49,46,47,47 | OK | 30 - 40 | 50,45,43,44,44 | OK | 20 - 30 | ||
| QET B/Ret | 35 | 29, 33, 47, 48 | 48,48,52,48 | OK | 45,45,49,45 | OK | 20 - 30 | ||||
| QET C/Ret | 7 | 5, 9, 10, 28 | 48,46,35,45 | OK | 45,43,33,42 | OK | 20 - 30 | ||||
| QET D/Ret | 38 | 36, 40, 43, 44 | 46,46,45,45 | OK | 43,43,42,43 | OK | 20 - 30 | ||||
| Charge | |||||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.036 | OK | 1.0 - 1.1 k | 1.042 | OK | 1.0 - 1.1 k | ||
| Qi bias/QET Ret | 15 | 24 - 26 nS | 40 M | OK | < 1 nS | OK | OK | <1 nS | |||
| Qo bias/QET Ret | 11 | 39 M | OK | OK | OK | <1 nS | |||||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 48 M, 48 M | OK | 6 - 8 nS | 50 M, 49 M | OK | <1 nS | ||
| Qo feedback/Drain/Source | 30 | 12, 27 | 48 M, 48 M | OK | 50 M, 49 M | OK | <1 nS | ||||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK | <0.1 nS | – | |
| Qo drain | 12 | chassis | – | OK | – | OK | <0.1 nS | – | |||
| Qi source | 31 | chassis | – | OK | – | OK | <0.1 nS | – | |||
| Qo source | 27 | chassis | – | OK | – | OK | <0.1 nS | – | |||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 1.03 | OK | 0.9 - 1.0 V | 1.089 | OK | — | ||
| LED 1 | chassis | 42 | 1.0 - 1.1 V | 1.231 | – | 1.2 - 1.5 V | 1.491 | – | — | – | |
| LED 2 | chassis | 26 | 1.221 | – | 1.497 | – | — | – | |||
| Cooldown checkout sheet (rev C SQUET) | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Detector: | G8 | BH: | Stripline: | 20 | |||||||
| SC: | ZB1-5-1 | 220 K | LN | LHe | |||||||
| Face: | 6 | Date | 2011-01-25 | Date | 2011-01-26 | Date | |||||
| SQUET: | ZC70 | Who | HC | Who | HC | Who | |||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | (Gnd lead) | |||||||||
| Nominal: | Nominal: | Nominal: | |||||||||
| < 0.1 nS | < 0.1 nS | < 0.1 nS | |||||||||
| Phonon | |||||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 3 | OK | 1 - 2 k | 2 | OK | 80 - 140 | ||
| SQUID bias B/Ret | 19 | 34 | 2.6 | OK | 1.7 | OK | 80 - 140 | ||||
| SQUID bias C/Ret | 8 | 24 | 2.9 | OK | 1.8 | OK | 80 - 140 | ||||
| SQUID bias D/Ret | 23 | 39 | 3.6 | OK | 2.1 | OK | 80 - 140 | ||||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 20 | OK | 9 - 13 k | 15 | OK | 1.0 - 1.3 k | ||
| Feedback B/Ret | 2 | 18 | 15 | OK | 9 | OK | 1.0 - 1.3 k | ||||
| Feedback C/Ret | 41 | 25 | 20 | OK | 14 | OK | 1.0 - 1.3 k | ||||
| Feedback D/Ret | 22 | 6 | 24 | OK | 17 | OK | 1.0 - 1.3 k | ||||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 52,47,45,45,45 | OK | 30 - 40 | 49,44,41,42,42 | OK | 20 - 30 | ||
| QET B/Ret | 35 | 29, 33, 47, 48 | 42,41,45,41 | OK | 38,37,41,37 | OK | 20 - 30 | ||||
| QET C/Ret | 7 | 5, 9, 10, 28 | 44,42,35,42 | OK | 41,39,32,39 | OK | 20 - 30 | ||||
| QET D/Ret | 38 | 36, 40, 43, 44 | 46,46,45,45 | OK | 43,42,42,42 | OK | 20 - 30 | ||||
| Charge | |||||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.034 | OK | 1.0 - 1.1 k | 1.039 | OK | 1.0 - 1.1 k | ||
| Qi bias/QET Ret | 15 | 24 - 26 nS | 39 M | OK | < 1 nS | OK | OK | <1 nS | |||
| Qo bias/QET Ret | 11 | 39 M | OK | OK | OK | <1 nS | |||||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 49 M, 49 M | OK | 6 - 8 nS | 50 M, 50 M | OK | <1 nS | ||
| Qo feedback/Drain/Source | 30 | 12, 27 | 49 M, 49 M | OK | 50 M, 50 M | OK | <1 nS | ||||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK | <0.1 nS | – | |
| Qo drain | 12 | chassis | – | OK | – | OK | <0.1 nS | – | |||
| Qi source | 31 | chassis | – | OK | – | OK | <0.1 nS | – | |||
| Qo source | 27 | chassis | – | OK | – | OK | <0.1 nS | – | |||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 1.026 | OK | 0.9 - 1.0 V | 1.081 | OK | — | ||
| LED 1 | chassis | 42 | 1.0 - 1.1 V | 1.265 | – | 1.2 - 1.5 V | 1.904 | – | — | – | |
| LED 2 | chassis | 26 | 1.26 | – | 1.958 | – | — | – | |||
| 2030 | Stop LN2 transfer |
|---|---|
| 2100 | Fill cold trap full |
| 2130 | Cryoclean mixture overnight: G1 = 492 mbar; G2 = 514 mbar; V6 open 13.0% |