| 0126 | Series 07101222_2351 stopped at 2nd file. Start a new series 07101223_0122 with the same settings. |
| 0149 | Series 07101223_0122 hung at the 2nd file event 26848. Start series 07101223_0150 |
| 0240 | Series 07101223_0150 hung again because trigger FIFO always read count 0. |
| 0241 | Started series 07101223_0239 and display 100% of the traces to make vi run slower. Trigger threshold: DCRC30 G8 = 0xD ADC bins; DCRC33 G9=0x14 ADC bins. |
| | A few debug runs |
| | 07101223_0401 with neutron source Trigger threshold: DCRC30 G8=3.5mV (0x17); DCRC33 G9=9mV (0x3b). vi stopped at file 5. |
| 0500 | Take background run 07101223_0503 after relocking squids: Trigger threshold: DCRC30 G8=3.5mV (0x17); DCRC33 G9=9mV (0x3b). |
| 0540 | Series 07101223_0503 stopped at file 5 |
| 0854 | Relocking phonon channels on DCRC33 G9. Start neutron shield background run 07101223_0853. |
| 1042 | Stop the series 07101223_0853 at file 6. |
| 1053 | Start the neutron run with shield on 07101223_1052 |
| 1218 | Stop 07101223_1052 at file 6 |
| 1225 | Start 07101223_1224 Ba-133 + Am241 run. |
| 1252 | 07101223_1224 stopped at file 4 |
| 1256 | Start 07101223_1256 same settings as 07101223_1224 |
| 1405 | Stop the run 07101223_1224. Lost one channel. |
| 1424 | Set QET bias =-1000uA to all phonon channels and charge bias=0 and take noise psd from labview. Other phonon settings are the same as data taking runs. |
| 1653 | 1K pot is out of liquid helium. Stop 1K pot and circulation and warm up. |
| | DCRC30 Qi charge noise is larger. |