| 1236 | Stop flashing the detectors. The tower temperature is around 870 during flashing. |
| 1241 | The oscillation noise shows up in charge channels. Power off both boards. |
| 1244 | Power on DCRC30 G9, there is still low frequency oscillation in charge, but not as big. Grounding the power supply helps. |
| 1246 | Power on DCRC33. The big noise come back again. |
| 1311 | DCRC33 itself has this huge noise. This seems can be reduced by grounding the board power. |
| 1549 | Removed the phonon base line oscillation by synchronizing the clock. |
| 2108 | Taking series 07101222_2105. Trigger on phonons. Trigger threshold: DCRC30 G8 3mV; DCRC33 G9 8mV. DAQ setting and Trigger setting |
| 2130 | Found series 07101222_2105 hung at fine number 3. |
| 2218 | Fill N2 cold trap full |
| 2311 | Stop the series 07101222_2138 at file 11, background run. PhD of one detector railed during the run. DCRC setting and Trigger settings. Trigger threshold: DCRC30 G8 3mV; DCRC33 G9 8mV. |
| 2321 | All channels on DCRC30 G8 are OK |
| 2325 | A and B of DCRC33 G9 railed. |
| 2329 | Fixed A and B on DCRC33 G9 by relocking the squids |
| 2347 | Put the neutron source ~40“ away from the magnetic shield, ~36” above the ground. |
| 2351 | Start 07101222_2351 with the neutron source. DCRC setting and Trigger settings. Trigger threshold: DCRC30 G8 3mV; DCRC33 G9 8mV. |