| 0932 | After overnight pumping on the main bath, leak checker P2=1.0E-2mbar, leak rate=1.3E-2cc/s. It's unlikely we can reach 1E-6cc/s in a reasonable amount of time. Cryocleaning mixture is still running. |
| 1343 | Vent the main bath through the leak checker. Got the physics leak checker to do a sniff test. |
| 1404 | Fill N2 cold trap full |
| 1518 | Found a suspicious leak at the middle In seal at BG 3.6E-5cc/s. Leak rate rose to ~4E-5cc/s |
| 1532 | Found a suspicious leak at the bottom of the baffle at BG 3.6E-5cc/s. Leak rate rose to ~8E-5cc/s |