2010/11/19

1005 The temperature shooting up while flashing yesterday is found due to using the wrong calibration curve for the tower thermometer. Using the flash settings 0.4mA, on 100us, and off 2500us, two-LED flashing heats the tower to ~650mK and four-LED flashing heats the tower to ~850mK.
1132 Exit all Labview vi's and upgrade Labview to version 2010
1446 To turn on DCRC clock: wr 15 4, wr 15 24. Test signal clock: wr 3c 2000
After replacing a few resistors on one board, there is no sign of noise improvement