1217 Pump out exchange gas, pump mixture into dump. MC T=78.1K
1339 Throughput test at 77K 16mbar/min, about half of the value in the past. No leak at background level 4.3E-9 cc/s
1356 Do the throughput test again: 17mbar/min. Still too low. Pump mixture out. Probably the mixture was not cleaned well. We'll go ahead and blow out liquid N2 and clean the mixture at the same time and redo the throughput test after that.
1411 G1 always goes up to ~100mbar after V5 is closed. It doesn't seen V12B is leaking. Not sure what happened.
1419 Start to blow out liquid N2. He background 6.4E-9 cc/s
Cryoclean mixture
1530 Liquid N2 is out. No gross leak found.
1550 Previous settings may not work. Try new settings and redo the cleaning
Cryocleanning settings
1659 Stop cryocleaning mixture, pump mixture into dump.
1732 Redo the throughput test after cleaning the mixture: 30mbar/min. The mixture was dirty. Pump out mixture.
1734 Pump the main bath.
1745 Leak level drops from ~1E-6cc/s to 3E-8cc/s when pumping the main bath. There is likely a leak.
1746 P2 reaches 1 mbar. Stop pumping bath and fill in He again
1820 Filled bath with He gas. Leak: 1.6E-6cc/s
Put in 0.1mbar He-4 exchange gas into IVC. (1mbar in exchange gas manifold → 0.063 mbar in IVC)
1852 Pump transfer stick
1917 Pump He transfer stick to 5E-3cc/s
Transfer liquid He to 35.1%. Start pumping exchange gas at 20K, MC temperature starts to rise from 15K. Pump the exchange gas overnight