2010/10/17

1200 SQUET card ZC-23 connected to DIB1
Electrical check conducted - all looks good
Cooldown checkout sheet (rev C SQUET)
Detector: G41 BH:
SC: ZB1-1-9 RT
Face: 8 Date 2010-10-17
SQUET: ZC23 Who HC
Signal Pos. lead Gnd lead Nominal Actual Chassis
(Gnd lead)
Nominal:
< 0.1 nS
Phonon
SQUID bias A/Ret 4 20 3 - 5 k 4.4 OK
SQUID bias B/Ret 19 34 5 OK
SQUID bias C/Ret 8 24 4.6 OK
SQUID bias D/Ret 23 39 4.3 OK
Feedback A/Ret 37 21 15 - 25 k 18 OK
Feedback B/Ret 2 18 20 OK
Feedback C/Ret 41 25 19 OK
Feedback D/Ret 22 6 18 OK
QET A/Ret 3 1, 13, 14, 17, 32 60 - 80 89,91,85,85,85 OK
QET B/Ret 35 29, 33, 47, 48 87,86,95,87 OK
QET C/Ret 7 5, 9, 10, 28 86,85,74,84 OK
QET D/Ret 38 36, 40, 43, 44 84,84,83,84 OK
Charge
FET heater +/- 45 46 1.0 - 1.1 k 1.068 OK
Qi bias/QET Ret 15 24 - 26 nS 26 OK
Qo bias/QET Ret 11 26 OK
Qi feedback/Drain/Source 50 16, 31 15 - 17 nS 23,23 OK
Qo feedback/Drain/Source 30 12, 27 23,23 OK
Qi drain 16 chassis < 0.1 nS – OK
Qo drain 12 chassis – OK
Qi source 31 chassis – OK
Qo source 27 chassis – OK
FET Tsense +/FET heater- 49 46 0.6 - 0.7 V 0.723 OK
LED 1 42 chassis 1.0 - 1.1 V 1.137 –
LED 2 26 chassis 1.127 –
1400 IVC sealed
Note: Remember to use vacuum grease! It took far too long to clean the flanges!
IVC evacuated
1549 Start He-3 and He-4 pump. Pump fridge from both ends. Open V9, V6, V5, and V1. Power glitch always turns off the pump. We may need something to prevent this. G2=652mbar.
1945 Background He leak rate 1E-9 mbar l/s
He leak check reveals leak rate of 1E-9 mbar l/s