2010/10/15
| 1400 | Conduct leak check |
|---|---|
| Test of internal leak shows no need for calibration | |
| Background He leak rate of 2.5E-9 mbar l/s | |
| He leak rate of 4.5E-9 mbar l/s, within acceptable parameters | |
| 1600 | Electrical check |
| QET A/Ret on stripline 20 is open | |
| Possible bad connection or defective SQUET card |
| Cooldown checkout sheet (rev C SQUET) | |||||
|---|---|---|---|---|---|
| Detector: | G41 | BH: | |||
| SC: | ZB1-1-9 | RT | |||
| Face: | 8 | Date | 2010-10-15 | ||
| SQUET: | ZC23 | Who | HC | ||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis |
| (Gnd lead) | |||||
| Nominal: | |||||
| < 0.1 nS | |||||
| Phonon | |||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.4 | OK |
| SQUID bias B/Ret | 19 | 34 | 5 | OK | |
| SQUID bias C/Ret | 8 | 24 | 4.6 | OK | |
| SQUID bias D/Ret | 23 | 39 | 4.3 | OK | |
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 18 | OK |
| Feedback B/Ret | 2 | 18 | 20 | OK | |
| Feedback C/Ret | 41 | 25 | 19 | OK | |
| Feedback D/Ret | 22 | 6 | 18 | OK | |
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 89,85,85,86,86 | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 86,86,95,86 | OK | |
| QET C/Ret | 7 | 5, 9, 10, 28 | 86,84,74,84 | OK | |
| QET D/Ret | 38 | 36, 40, 43, 44 | 84,83,83,84 | OK | |
| Charge | |||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.068 | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | 26 | OK | |
| Qo bias/QET Ret | 11 | 26 | OK | ||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 23,23 | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 23,23 | OK | |
| Qi drain | 16 | chassis | < 0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | |
| Qi source | 31 | chassis | – | OK | |
| Qo source | 27 | chassis | – | OK | |
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.725 | OK |
| LED 1 | 42 | chassis | 1.0 - 1.1 V | 1.137 | – |
| LED 2 | 26 | chassis | 1.128 | – | |
| Cooldown checkout sheet (rev C SQUET) | |||||
|---|---|---|---|---|---|
| Detector: | G41 | BH: | |||
| SC: | ZB1-1-6 | RT | |||
| Face: | 5 | Date | 2010-10-15 | ||
| SQUET: | ZC15 | Who | HC | ||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis |
| (Gnd lead) | |||||
| Nominal: | |||||
| < 0.1 nS | |||||
| Phonon | |||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 5 | OK |
| SQUID bias B/Ret | 19 | 34 | 4.6 | OK | |
| SQUID bias C/Ret | 8 | 24 | 4.2 | OK | |
| SQUID bias D/Ret | 23 | 39 | 4.1 | OK | |
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 21 | OK |
| Feedback B/Ret | 2 | 18 | 19 | OK | |
| Feedback C/Ret | 41 | 25 | 18 | OK | |
| Feedback D/Ret | 22 | 6 | 17 | OK | |
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | all open | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 72,72,81,72 | OK | |
| QET C/Ret | 7 | 5, 9, 10, 28 | 70,69,61,68 | OK | |
| QET D/Ret | 38 | 36, 40, 43, 44 | 70,69,69,69 | OK | |
| Charge | |||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.063 | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | open for A, else 26 | OK | |
| Qo bias/QET Ret | 11 | open for A, else 26 | OK | ||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 23,23 | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 24,23 | OK | |
| Qi drain | 16 | chassis | < 0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | |
| Qi source | 31 | chassis | – | OK | |
| Qo source | 27 | chassis | – | OK | |
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.71 | OK |
| LED 1 | 42 | chassis | 1.0 - 1.1 V | 1.134 | – |
| LED 2 | 26 | chassis | 1.125 | – | |
| 2100 | Leave IVC to pump overnight to clean any contamination |
|---|