2010/10/13

1810 Detector level continuity checks on G41
Work conducted on clean bench

Detector Level Continuity Checks

Detector G41 DIB1
date/who: 2010/10/13 HC
+lead - lead signal typical* measured
1 2 QET A 100 to 200 58 Ohms
1 3 QETA-B 100 to 200 195 Ohms
3 4 QET B 100 to 200 196 Ohms
3 5 QETB-C 100 to 200 269 Ohms
5 6 QET C 100 to 200 167 Ohms
5 7 QETC-D 100 to 200 229 Ohms
7 8 QET D 100 to 200 146 Ohms
7 1 QETD-A 100 to 200 152 Ohms
9 chassis LED1 Rforward 9M 12 MOhms
chassis 9 LED1 Rreverse <1nS OK
10 chassis LED2 Rforward 9M 12 MOhms
chassis 10 LED2 Rreverse <1nS OK
11 12 Qo-Qi 125 to 175 162 Ohms
1 11 Qi-QETs 50 to 100 100 Ohms
1 12 Qo-QETs 100 to 200 175 Ohms
Short to chassis All signal except LEDs <1nS OK

* mZIP values

Detector G41 DIB2
date/who: 2010/10/13 HC
+lead - lead signal typical* measured
1 2 QET A 100 to 200 51 Ohms
1 3 QETA-B 100 to 200 134 Ohms
3 4 QET B 100 to 200 55 Ohms
3 5 QETB-C 100 to 200 190 Ohms
5 6 QET C 100 to 200 162 Ohms
5 7 QETC-D 100 to 200 210 Ohms
7 8 QET D 100 to 200 72 Ohms
7 1 QETD-A 100 to 200 161 Ohms
9 chassis LED1 Rforward 9M 12 MOhms
chassis 9 LED1 Rreverse <1nS OK
10 chassis LED2 Rforward 9M 12 MOhms
chassis 10 LED2 Rreverse <1nS OK
11 12 Qo-Qi 125 to 175 153 Ohms
1 11 Qi-QETs 50 to 100 108 Ohms
1 12 Qo-QETs 100 to 200 180 Ohms
Short to chassis All signal except LEDs <1nS OK

* mZIP values

1845 Detector G41 placed on tower
Am-241 source placed above B1
Detector and source holder clamped together and position 1 side coaxes attached
Tower level continuity checks conducted
Work conducted on clean bench

Tower Level Continuity Checks

Detector G41 DIB1
side coax 1 (ZB1-1-6) date/who: 2010/10/13 HC
face
+lead - lead signal typical measured
1 2 QET A 100 to 200 143 Ohms
1 3 QETA-B 100 to 200 274 Ohms
3 4 QET B 100 to 200 278 Ohms
3 5 QETB-C 100 to 200 346 Ohms
5 6 QET C 100 to 200 250 Ohms
5 7 QETC-D 100 to 200 308 Ohms
7 8 QET D 100 to 200 231 Ohms
7 1 QETD-A 100 to 200 234 Ohms
9 chassis LED1 Rforward 9M 12 MOhms
chassis 9 LED1 Rreverse <1nS OK
10 chassis LED2 Rforward 9M 12 MOhms
chassis 10 LED2 Rreverse <1nS OK
1 11 Qo Rbias 24nS to 26nS 20 nS
11 12 Qo C coupling <1nS OK
12 13 Qo Rfeedback 24nS to 26nS 21 nS
1 13 <1nS OK
1 14 24nS to 26nS 21 nS
14 15 Qi C coupling <1nS OK
15 16 Qi Rfeedback 24nS to 26nS 21 nS
Short to chassis All signal except LEDs <1nS OK
Detector G41 DIB2
side coax 1 (ZB1-1-9) date/who: 2010/10/13 HC
face
+lead - lead signal typical measured
1 2 QET A 100 to 200 140 Ohms
1 3 QETA-B 100 to 200 220 Ohms
3 4 QET B 100 to 200 145 Ohms
3 5 QETB-C 100 to 200 276 Ohms
5 6 QET C 100 to 200 252 Ohms
5 7 QETC-D 100 to 200 297 Ohms
7 8 QET D 100 to 200 163 Ohms
7 1 QETD-A 100 to 200 248 Ohms
9 chassis LED1 Rforward 9M 12 MOhms
chassis 9 LED1 Rreverse <1nS OK
10 chassis LED2 Rforward 9M 12 MOhms
chassis 10 LED2 Rreverse <1nS OK
1 11 Qo Rbias 24nS to 26nS 21 nS
11 12 Qo C coupling <1nS OK
12 13 Qo Rfeedback 24nS to 26nS 21 nS
1 13 <1nS OK
1 14 24nS to 26nS 21 nS
14 15 Qi C coupling <1nS OK
15 16 Qi Rfeedback 24nS to 26nS 21 nS
Short to chassis All signal except LEDs <1nS OK
2040 Assembled tower placed in N2 purge cabinet, bottom shelf