start › cdms › k100 › run_diary › 20100722

2010-07-22

Tower resistance check

This is the old table, need to type in new values [image not archived: fixme.svg]

Detector G9
side coax Z13 1-6 date/who: 2010-02-11 JZ
face 5
+lead - lead signal typical measured
1 2 QET A 100 to 200 107.5
2 4 QETA-B 100 to 200 115.5
3 4 QET B 100 to 200 108.3
4 6 QETB-C 100 to 200 114.6
5 6 QET C 100 to 200 109.8
6 8 QETC-D 100 to 200 118.0
7 8 QET D 100 to 200 115.2
2 8 QETA-D 100 to 200 125.6
chassis 9 LED1 Rforward 8M to 9M 1.49
chassis 10 LED2 Rforward 8M to 9M 1.48
chassis 9 LED1 Vforward 1.0V to 1.1V 1.102
chassis 10 LED2 Vforward 1.0V to 1.1V 1.100
9 chassis LED1 Rreverse <1nS OK
10 chassis LED2 Rreverse <1nS OK
1 11 Qo Rbias 24nS to 26nS 38.1MOhm
12 11 Qo C coupling <1nS OK
12 13 Qo Rfeedback 24nS to 26nS 38.3MOhm
1 14 Qi Rbias 24nS to 26nS 38.4MOhm
15 14 Qi C coupling <1nS OK
15 16 Qi Rfeedback 24nS to 26nS 38.5MOhm

Cooldown resistance check

Cooldown checkout sheet (rev C SQUET)
Detector: G9 BH:
SC: RT LN
Face: Date 2010-07-22 Date 2010-07-25
SQUET: ZC15 Who HC & JZ Who JZ
Signal Pos. lead Gnd lead Nominal Actual Chassis Nominal Actual Chassis
(Gnd lead) (Gnd lead)
Nominal: Nominal:
< 0.1 nS < 0.1 nS
Phonon
SQUID bias A/Ret 4 20 3 - 5 k 5 OK 1 - 2 k 2 OK
SQUID bias B/Ret 19 34 4.6 OK 1.7 OK
SQUID bias C/Ret 8 24 4.2 OK 1.5 OK
SQUID bias D/Ret 23 39 4.1 Ok 1.5 OK
Feedback A/Ret 37 21 15 - 25 k 21 OK 9 - 13 k 8.4 OK
Feedback B/Ret 2 18 19 Ok 7.4 OK
Feedback C/Ret 41 25 18 OK 7 OK
Feedback D/Ret 22 6 17.5 OK 6.8 OK
QET A/Ret 3 1, 13, 14, 17, 32 60 - 80 87,73,71,71,70 OK 30 - 40 50,39,36,37,36 OK
QET B/Ret 35 29, 33, 47, 48 72,71,75,72 OK 37,37,40,37 OK
QET C/Ret 7 5, 9, 10, 28 73,68,61,68 OK 37,34,27,34 OK
QET D/Ret 38 36, 40, 43, 44 69,69,68,69 OK 36,36,36,36 OK
Charge
FET heater +/- 45 46 1.0 - 1.1 k 1.063 k OK 1.0 - 1.1 k 1.054 OK
Qi bias/QET Ret 15 24 - 26 nS 39 M OK < 1 nS OK OK
Qo bias/QET Ret 11 39 M OK OK OK
Qi feedback/Drain/Source 50 16, 31 15 - 17 nS OK OK 6 - 8 nS OK OK
Qo feedback/Drain/Source 30 12, 27 OK OK OK OK
Qi drain 16 chassis < 0.1 nS – OK < 0.1 nS – OK
Qo drain 12 chassis – OK – OK
Qi source 31 chassis – OK – OK
Qo source 27 chassis – OK – OK
FET Tsense +/FET heater- 49 46 0.6 - 0.7 V 0.7 OK 0.9 - 1.0 V 1.082 OK
LED 1 chassis 42 1.0 - 1.1 V 1.1 – 1.2 - 1.5 V 1.483 –
LED 2 chassis 26 1.1 – 1.492 –