2010/06/11
| 11:15 | Resistance check at 77K: OK |
|---|
| Cooldown checkout sheet (rev C SQUET) | ||||||||
|---|---|---|---|---|---|---|---|---|
| Detector: | No detector | BH: | ||||||
| SC: | ?? | RT | LN | |||||
| Face: | 4 | Date | Date | 2010-06-11 | ||||
| SQUET: | ZC15 | Who | Who | JZ | ||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | |||||||
| Nominal: | Nominal: | |||||||
| < 0.1 nS | < 0.1 nS | |||||||
| Phonon | ||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 1 - 2 k | 2 | OK | ||
| SQUID bias B/Ret | 19 | 34 | 1.7 | OK | ||||
| SQUID bias C/Ret | 8 | 24 | 1.5 | OK | ||||
| SQUID bias D/Ret | 23 | 39 | 1.5 | OK | ||||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 9 - 13 k | 8.4 | OK | ||
| Feedback B/Ret | 2 | 18 | 7.4 | OK | ||||
| Feedback C/Ret | 41 | 25 | 7 | OK | ||||
| Feedback D/Ret | 22 | 6 | 6.8 | OK | ||||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 30 - 40 | 47.3, 35.8, 34.3, 34.9, 34.8 | OK | ||
| QET B/Ret | 35 | 29, 33, 47, 48 | 35.4, 35.1, 37.2, 35.2 | OK | ||||
| QET C/Ret | 7 | 5, 9, 10, 28 | 33.6, 33.4, 26.2, 33.2 | OK | ||||
| QET D/Ret | 38 | 36, 40, 43, 44 | 35, 34.6, 34.3, 34.4 | OK | ||||
| Charge | ||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.0 - 1.1 k | 1.052 | OK | ||
| Qi bias/QET Ret | 15 | 24 - 26 nS | < 1 nS | OK | OK | |||
| Qo bias/QET Ret | 11 | OK | OK | |||||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 6 - 8 nS | 48.1MOhm, 48.1MOhm | OK | ||
| Qo feedback/Drain/Source | 30 | 12, 27 | 48.5MOhm, 48.5MOhm | OK | ||||
| Qi drain | 16 | chassis | < 0.1 nS | – | < 0.1 nS | – | OK | |
| Qo drain | 12 | chassis | – | – | OK | |||
| Qi source | 31 | chassis | – | – | OK | |||
| Qo source | 27 | chassis | – | – | OK | |||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.9 - 1.0 V | 1.081 | OK | ||
| LED 1 | chassis | 42 | 1.0 - 1.1 V | – | 1.2 - 1.5 V | OK | – | |
| LED 2 | chassis | 26 | – | OK | – | |||