UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20100303

2010/03/03

Run Diary

Reset DCRC

Time
10:40 Save the following settings as Preset 2
SQUID settings
Channel Frontend Gain Output Gain Feedback Null Driver Offset SQUID Bias (uA) Lockpoint (mV) QET Bias(uA)
A 7 1 0 0 95 3.6 -136.9
B 7 1 0 0 100 3.4 -120.1
C 7 1 0 0 92 3.2 -120.1
D 7 1 0 0 95 3.5 -120.1
Q Inner (V) 3
Q Outer (V) 3
Time
10:45 Soft reset DCRC and then load Preset 2
Open I:\Labview\DCRC\20100303\DCRC.llb\SCDMS_DetectorReadout.vi and checked that the above operation worked.
Lookback time spikes still show up once in a while when triggering on phonon channels.
Pulses after loading Preset 2
Spikes show up once in a while when triggering on phonon channels

Test LED Flashing

Time
11:21 Tower T = 24.8mK (224.1kOhm). Try LED 1 only. I = 400 uA (164/4095); ON = 100 us; OFF = 800 us.
LED DC offset found not fixed
14:09 LED DC offset fixed in version 20100303
14:23 DAQ stopped for an error when need to flash
Error
Time
14:50 The same error occurred even only the readout vi was running