start › cdms › k100 › run_diary › 20091211

2009/12/11

DCRC Calibration Table

Quantity Decimal Hex Output
Qi Bias 0.04 201E -0.8 mV
Qo Bias -0.04 1FE3 -0.3 mV
SQUID Bias A 0.83 808 0.1 mV
SQUID Bias B 0.24 802 -0.3 mV
SQUID Bias C 1.51 80F -0.8 mV
SQUID Bias D -0.24 7FD -0.1 mV
QET Bias A 34.1 20BC 0.05 mV
QET Bias B -1.7 1FF6 0.8 mV
QET Bias C -1.2 1FF9 0.05 mV
QET Bias D -0.1
-0.3
1FFF
1FFE
0.26 mV
-0.26

Set the above settings and Frontend Gain = 0, Output Gain = 1, Feedback Null = 0, Driver Offset = 0, Openloop, and Preamp on all channels as the Preset 1.

There is a 1.5 kOhm resistor in the SQUID bias line. A 9 V battery can only supply 6 mA current at most. Zapping using the battery had no effect. Zapping A using DCRC only changed MC temperature from 39.2 mK to 56 mK. Battery output much more power but the results were worse than that of DCRC.