2009/12/08
Run diary
| Time | Task |
|---|---|
| 08:42 | Set Qi = 0.05 V (x2020) and Qo = -0.04 V (x1FE6) to zero the bias; Turn on FET heater. |
| 08:45 | Turn off heater at FET T = 0.893. Charge channels are on. MC temperature starts to go up from 33.5 mK. |
| 08:49 | East lights off. Fridge grounded to orange ground. MC T = 34.3 mK. Take noise traces of Qi and Qo: run001 and run002. |
| 08:51 | Ground fridge to Cu rod. MC T = 34.6 mK. Qi run003 |
| 08:52 | Qo run004 |
| 08:54 | Ground to black ground. MC T = 35.1 mK. Qi spd run005 |
| 08:55 | Qo psd run006 |
| plot of run001, 3, and 5 |
|---|
![]() |
| matlab .fig file |
| plot of run002, 4, and 6 |
![]() |
| matlab .fig file |
| All peaks overlap. There is no essential difference between different grounds. Low frequency base line difference are due to pulses. |
| Time | Task |
|---|---|
| 09:53 | Take Qi noise psd with an without LS370 connected. run007 and run008. |
| plot of run007 and run008 |
|---|
![]() |
| matlab .fig file |
| All peaks overlap. |
| Time | Task |
|---|---|
| 16:12 | Turn everything off and see the effect on charge noise psd. Take run009 to 011. |
| Plot of run009 to 011 |
|---|
![]() |
| matlab .fig file |
| The blue curve is a repetition of the green. Temperature may change between them since pumps are off. Can not see the effects of the electronics we can turn off. |
| Time | Task |
|---|---|
| 11:00 | Disconnect DCRC and solder wires on LED 1 and 2. |
| 11:46 | Plug in DCRC. |
| 13:09 | Tune SQUID, zero phonon and charge biases. |
| Channel | FE Gain | Output Gain | Feedback Null (mV) | Driver Offset (V) | SQUID Bias (uA) | Lockpoint (mV) | QET Bias (uA) |
|---|---|---|---|---|---|---|---|
| A | 0 | 1 | 0.02 | 0.02 | 90.01 | 3.20 | 36.9 |
| B | 0 | 1 | 0.02 | 0.00 | 80.93 | 2.40 | 0.6 |
| C | 0 | 1 | 0.02 | 0.00 | 76.34 | 2.38 | -0.1 |
| D | 0 | 1 | 0.02 | 0.02 | 69.01 | 1.72 | 1.5 |
| Qi (V) | 0.05 (x2020) |
|---|---|
| Qo (V) | -0.04 (x1EF6) |
| LED Current Settings |
|---|
| 4095 = Full range 10 mA. Polarity is negative on DCRC v1. |
| Time | Task |
|---|---|
| 13:22 | Close loop to non-inverting. Set FE gain = 1 for all channels and turn off test signal. |
| 13:26 | Start to flash the LED's. First try single shot mode and monitor the output. |
| 13:31 | Try LED 1 only. I = 400 uA (164/4095); ON = 50 us; OFF = 2500 us. MC T = 33.3 mK. |
| 13:37 | Flash continuously. Turn LED 1 on and off, no change on the bias voltage. |
| 15:22 | Turn on flashing and wait for a few minutes, fridge temperature doesn't change. Suspect flashing doesn't work. |
| 16:31 | Tried the newest version of detector control vi. Flashing doesn't work either. |
Data List
All data can be found at http://www.hep.umn.edu/cdms/cdms_restricted/K100/runs/run001/20091208/
| Charge channel noise PSD |
|---|
| East lights off. MC T ~ 35 mK. LS370 connected and on. |
| Bias (V) | Hex | |
| Qi | 0.05 | 2020 |
| Qo | -0.04 | 1FE6 |
| Time | Run # | Comment |
|---|---|---|
| 1 | Fridge grounded to orange ground; Qi noise PSD. | |
| 2 | Fridge grounded to orange ground; Qo noise PSD. | |
| 3 | Fridge grounded to Cu rod; Qi noise PSD. | |
| 4 | Fridge grounded to Cu rod; Qo noise PSD. | |
| 5 | Fridge grounded to black ground; Qi noise PSD. | |
| 6 | Fridge grounded to black ground; Qo noise PSD. |
| MC T = 36 mK. |
| Time | Run # | Comment |
|---|---|---|
| 09:53 | 7 | Fridge grounded to orange ground; Qi noise PSD. LS370 connected. |
| 8 | Fridge grounded to orange ground; Qi noise PSD. LS370 disconnected. |
| Time | Run # | Comment |
|---|---|---|
| 16:12 | 9 | Fridge grounded to orange ground; Qi noise PSD. LS370 connected. East lights off. |
| 10 | Fridge grounded to orange ground; Qi noise PSD. LS370 disconnected. IGH, ILM, pumps, cleamroom off. | |
| 11 | Repeat run010. |