11/20/2009 9:47:27 AM

MC temperature fluctuates.

H35_20091120_0946.PNG

igh_20091120_0946.PNG

11/20/2009 2:28:02 PM

Take data set Run001 to Run003

Can see the same noise peaks in the psd taken by the scope at the feedback line.

run001_to_002.png

11/20/2009 2:57:50 PM

Fill N2 cold trap full.

11/20/2009 3:34:45 PM

Take traces Run004 and Run005

Compare spectrum at QET bias line and the DCRC. It seems the input line is already contaminated by noise.

run004_to_005.png

11/20/2009 4:00:02 PM

Set QETb bias to 1480 uA to drive it normal and compare the noise at the input and output sides.

11/20/2009 4:13:37 PM

Take psd run006 and run007

compare psd at normal sate

run006_to_007.png

11/20/2009 4:22:33 PM

Set QETB=0. When QETB=1480 uA, oscilloscope DC voltage = 20.4 mV.

11/20/2009 5:27:05 PM

Compare AC and DC methods of measuring the Ic at a given temperature. Send a triangle wave down the QET biasing line, monitor both the QET bias and the feedback line.

At T = (36.75 +/- 0.25) mK, 57 Hz the Ic's are:

85 mVpp + 0 mVDC or 10 mVpp + 17 mVDC (10mVpp is the smallest signal that can show up in the scope in this measurement)

There is quite a difference between two different methods. This could be an AC effect or temperature effect or both.

11/20/2009 6:38:48 PM

Fill N2 cold trap full.

Data list

Psd

Time Run # MC T (mK) Channel FB Gain Output Gain SQUID Bias (uA) Lock Point (mV) QET Bias (uA) Qi Bias (V) Qo Bias (V) Comment
11/20/2009 2:32:08 PM 1  34.5  B 7 1 95 3.4 0 0 0  TES SC, Fridge grounded to Orange ground. Scope monitor FB
   2  34.5  B  7  1  95 3.4 0  0 0 Scope PSD
  3  34.5  B  7 1  95 3.4 0  0 0 DCRC PSD with Scope by battery. The inputs of scope are actually ioslated.
 11/20/2009 3:30:39 PM  4 35.4  B               Scope, QET bias
 11/20/2009 3:34:00 PM  5  35.4 B                DCRC
 11/20/2009 4:14:33 PM  6             1480     DCRC
  7              1480     Scope, QET bias