Cooldown checkout summary
Run 8
| Cooldown checkout sheet (rev C SQUET) |
| Detector: | G9 | BH: | | Stripline: | 19 | |
| SC: | ZB1-4-6 | RT | LN | LHe |
| Face: | 2 | Date | 2010-12-10 | Date | 2010-12-16 | Date | |
| SQUET: | ZC23 | Who | HC&JZ | Who | HC | Who | HC |
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| | | | | | (Gnd lead) | | | (Gnd lead) | | | (Gnd lead) |
| | | | | | Nominal: | | | Nominal: | | | Nominal: |
| | | | | | < 0.1 nS | | | < 0.1 nS | | | < 0.1 nS |
| Phonon |
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.4 | OK | 1 - 2 k | 1.6 | OK | 80 - 140 | | |
| SQUID bias B/Ret | 19 | 34 | | 5 | OK | | 2 | OK | 80 - 140 | | |
| SQUID bias C/Ret | 8 | 24 | | 4.6 | OK | | 1.7 | OK | 80 - 140 | | |
| SQUID bias D/Ret | 23 | 39 | | 4.3 | OK | | 1.6 | OK | 80 - 140 | | |
| | | | | | | | | | | | |
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 19 | OK | 9 - 13 k | 7.3 | OK | 1.0 - 1.3 k | | |
| Feedback B/Ret | 2 | 18 | | 20 | OK | | 8.2 | OK | 1.0 - 1.3 k | | |
| Feedback C/Ret | 41 | 25 | | 19 | OK | | 7.6 | OK | 1.0 - 1.3 k | | |
| Feedback D/Ret | 22 | 6 | | 18 | OK | | 7.1 | OK | 1.0 - 1.3 k | | |
| | | | | | | | | | | | |
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 91,100,85,86,86 | OK | 30 - 40 | 50,47,45,45,45 | OK | 20 - 30 | | |
| QET B/Ret | 35 | 29, 33, 47, 48 | | 87,87,97,87 | OK | | 46,45,48,45 | OK | 20 - 30 | | |
| QET C/Ret | 7 | 5, 9, 10, 28 | | 86,85,74,84 | OK | | 45,44,33,43 | OK | 20 - 30 | | |
| QET D/Ret | 38 | 36, 40, 43, 44 | | 84,84,84,84 | OK | | 44,43,43,43 | OK | 20 - 30 | | |
| | | | | | | | | | | | |
| Charge |
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.068 | OK | 1.0 - 1.1 k | 1.043 | OK | 1.0 - 1.1 k | | |
| | | | | | | | | | | | |
| Qi bias/QET Ret | 15 | | 24 - 26 nS | 25 | OK | < 1 nS | OK | OK | <1 nS | | |
| Qo bias/QET Ret | 11 | | | 26 | OK | | OK | OK | <1 nS | | |
| | | | | | | | | | | | |
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 23,23 | OK | 6 - 8 nS | 20,20 | OK | <1 nS | | |
| Qo feedback/Drain/Source | 30 | 12, 27 | | 23,23 | OK | | 20,20 | OK | <1 nS | | |
| | | | | | | | | | | | |
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK | <0.1 nS | – | |
| Qo drain | 12 | chassis | | – | OK | | – | OK | <0.1 nS | – | |
| Qi source | 31 | chassis | | – | OK | | – | OK | <0.1 nS | – | |
| Qo source | 27 | chassis | | – | OK | | – | OK | <0.1 nS | – | |
| | | | | | | | | | | | |
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.723 | OK | 0.9 - 1.0 V | 1.091 | OK | — | | |
| LED 1 | chassis | 42 | 1.0 - 1.1 V | 1.138 | – | 1.2 - 1.5 V | 1.49 | – | — | | – |
| LED 2 | chassis | 26 | | 1.126 | – | | 1.496 | – | — | | – |
| Cooldown checkout sheet (rev C SQUET) |
| Detector: | G8F | BH: | | Stripline: | 20 | |
| SC: | ZB1-5-1 | RT | LN | LHe |
| Face: | 6 | Date | 2010-12-10 | Date | 2010-12-16 | Date | |
| SQUET: | ZC70 | Who | HC & JZ | Who | HC | Who | |
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| | | | | | (Gnd lead) | | | (Gnd lead) | | | (Gnd lead) |
| | | | | | Nominal: | | | Nominal: | | | Nominal: |
| | | | | | < 0.1 nS | | | < 0.1 nS | | | < 0.1 nS |
| Phonon |
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.9 | OK | 1 - 2 k | 2 | OK | 80 - 140 | | |
| SQUID bias B/Ret | 19 | 34 | | 4.4 | OK | | 1.7 | OK | 80 - 140 | | |
| SQUID bias C/Ret | 8 | 24 | | 4.7 | OK | | 1.8 | OK | 80 - 140 | | |
| SQUID bias D/Ret | 23 | 39 | | 5.2 | OK | | 2.1 | OK | 80 - 140 | | |
| | | | | | | | | | | | |
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 29 | OK | 9 - 13 k | 15 | OK | 1.0 - 1.3 k | | |
| Feedback B/Ret | 2 | 18 | | 26 | OK | | 9 | OK | 1.0 - 1.3 k | | |
| Feedback C/Ret | 41 | 25 | | 28 | OK | | 14 | OK | 1.0 - 1.3 k | | |
| Feedback D/Ret | 22 | 6 | | 35 | OK | | 17 | OK | 1.0 - 1.3 k | | |
| | | | | | | | | | | | |
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 85,83,80,81,80 | OK | 30 - 40 | 48,46,43,43,43 | OK | 20 - 30 | | |
| QET B/Ret | 35 | 29, 33, 47, 48 | | 78,77,80,78 | OK | | 38,38,41,38 | OK | 20 - 30 | | |
| QET C/Ret | 7 | 5, 9, 10, 28 | | 78,76,69,76 | OK | | 42,40,32,39 | OK | 20 - 30 | | |
| QET D/Ret | 38 | 36, 40, 43, 44 | | 80,80,80,80 | OK | | 43,43,42,43 | OK | 20 - 30 | | |
| | | | | | | | | | | | |
| Charge |
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.061 | OK | 1.0 - 1.1 k | 1.04 | OK | 1.0 - 1.1 k | | |
| | | | | | | | | | | | |
| Qi bias/QET Ret | 15 | | 24 - 26 nS | 25 | OK | < 1 nS | OK | OK | <1 nS | | |
| Qo bias/QET Ret | 11 | | | 25 | OK | | OK | OK | <1 nS | | |
| | | | | | | | | | | | |
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 23,23 | OK | 6 - 8 nS | 20,20 | OK | <1 nS | | |
| Qo feedback/Drain/Source | 30 | 12, 27 | | 23,23 | OK | | 20,20 | OK | <1 nS | | |
| | | | | | | | | | | | |
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK | <0.1 nS | – | |
| Qo drain | 12 | chassis | | – | OK | | – | OK | <0.1 nS | – | |
| Qi source | 31 | chassis | | – | OK | | – | OK | <0.1 nS | – | |
| Qo source | 27 | chassis | | – | OK | | – | OK | <0.1 nS | – | |
| | | | | | | | | | | | |
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | open | OK | 0.9 - 1.0 V | open | OK | — | | |
| LED 1 | chassis | 42 | 1.0 - 1.1 V | 1.155 | – | 1.2 - 1.5 V | 1.895 | – | — | | – |
| LED 2 | chassis | 26 | | 1.149 | – | | 1.951 | – | — | | – |