Pump Mixture to Dump
- (Not necessary) Close all valves on IGH.
- Open V5 and V9.
- Open the valve(s) to the cold trap in case there is mixture left in there from being circulated through them (V12A/B and speedivalve on cold trap 1).
- Open V6 all the way. This will take 3-4 minutes.
- When P1<0.014 mbar, mixture has been pumped sufficiently to dump. The reading on G2 should be ~ 800 mbar, and G1 should be < 1 mbar.
- To close off the dump, close V5 and V9.