Pump IVC
You are going to evacuate the IVC now. First, the scroll pump will be used to get the pressure in the IVC to 5.0E-2 mbar, and then the turbo pump will be opened to the IVC to continue to lower the pressure to ~ 1.0E-5 mbar. You'll need to turn on the turbo pump ahead of time (while isolated) because it needs to warm up.
- Make sure the check valve to IVC is closed.
- Close all valves on the pumping manifold.
- Turn on pumping station. Rough pumping is now on.
- Push the far right button on the pumping manifold to start warming up the turbo pump.
- When pressure reaches ~ 8.6E-2 mbar, turbo pumping will begin. The turbo pump cannot handle atmospheric pressure; it's fine to use it, however, when the pressure gauge on the manifold reads ~8.6E-2 mbar or below.
- Plug in the scroll pump to a power outlet - this automatically turns it on.
- Open the two valves leading from scroll pump to IVC line - this does NOT include the valve that isolates the turbo pump. You are now pumping on the IVC with the scroll pump.
- Monitor the pressure reading on the pumping manifold. Once it reads ~10E-2 mbar and the turbo pump has warmed up entirely, you can open the valve leading to the turbo pump. Isolate the scroll pump.
- Be sure to note in the run diary when you started pumping out the IVC.
- Pumping out the IVC will take about 5 hours. The target pressure is < 3.0E-5 mbar. If the IVC has diffculty reaching that pressure in this time frame, that might suggest there is a leak. If the pressure reading stays at 4.0E-3 mbar, that means the gauge is stuck and can't read below that pressure; in this case, just pump out for 5 hours and move forward in the cooldown process.