Evacuate 1 K Pot
Now you are going to evacuate on the 1 K pot. Similar to the main bath (MB), you want to make sure there is no contamination left in there as you start to cool-down; otherwise, contaminants will freeze in the 1 K pot.
- Make sure the speedivalve connected to the 1K pot is closed (should have already been closed from previous steps, but check nonetheless).
- Make sure the He4 gas supply is off and the vent valve is closed.
- Turn on He4 pump.
- Open V5A, V2A, V4A (this pumps out leftover He4 gas in lines before we pump out 1K pot line).
- Fully open 1 K pot Speedivalve and then close it.
- Close V5A, V2A, V4A.