Tower slowly reached 75 mK this morning.
RevC Pulseviewer not working. I took two datasets without squid biased and one dataset with squid biased to see IbIs and noise.
Zapped SQUID:
Before:

After :

Shubham processed the IbIs dataset but it is showing sinusoidal pattern.
Before zapping:

After zapping:

IbIs curve after zapping with RevE:
1000 to 300 uA :

500 uA to 0 uA :

Noise with RevE:
The noise profile on other datasets need to be processed, but a manual look with RevE showing applying QET bias current leads to change in overall noise plateau.
Note: