| Teleconference link: | https://umn.zoom.us/j/9024497080?omn=95322888432 |
| Meeting ID: | 902 449 7080 |
| Domestic: | 1-408-638-0968 (US toll) |
| Access code: |
- Lab Update
- Tower temperature at 90-100 mK while MC stays flat at 30 mK.
- Unplugging both thermometry cables donot show any effect on tower temperature.
- Unplugged stripline cables (with MC cable connected to monitor temperature),shows a weak effect but tower still above 80 mK.
- Prof. Paul thinks that since MC temperature is flat and stable, it must be some mechanical touch.
- Priority Updates
- General Updates
- Himangshu
- Shubham
- Zach
- Elliott
- Scott