  • Lab Update
    • Electrical checks done at 77K. No shorts or open connections.
    • No leak showing up during LN2 blowout.
    • Throughput test still shows low mixture throughput < 10 mbar/min. Previous runs have throughput as high as 56 mbar/min at this stage.
    • Mixture was cryocleaned with last reading of <50 mbar contaminants.
    • Other suspect is He-trap. Removed trap from slop and warmed it with heat-gun. No visible change in pressure, read on G1(least count ~10 mbar).
    • I am continuing another cryoclean and check the throughput tomorrow. I will keep the mixture in the loop after cryocleaning to let it be at 77K.