- Lab Update
- Lowered MB again. Checking leaks on each indium seal shows leak level at 10^-7 mbar L/s.
- This leaks might be strong when MB is up.
- Further tightened all indium seals, metal cracking sound on every seal.
- Leak tesking shows leak in 10^-8 mbar L/s.
- Raised MB and pumped out MB. Added He gas and pumped out again. Level remain within 10^-8 mmar L/s.
- Proceeding with LN2 transfer tomorrow.
- Priority Updates
- General Updates
- Himangshu
- Completed First Remote Shift yesterday. Found a few alarms going off with tes-uoft server and not on monitoring node. Ry and Mooaz will look into the issue further.
- A few slides on DMC work.
- Shubham
- Zach
- Elliott
- Scott