Teleconference link: https://umn.zoom.us/j/9024497080?omn=95322888432
Meeting ID: 902 449 7080
Domestic: 1-408-638-0968 (US toll)
Access code:
  • Lab Update
    • Investigated the ground loop that is seen on the striplines further.
    • Unplugging and plugging back connections on the top of the fridge doesn't seem to help directly.
    • Unplugged all SQUETS and swapping SQUETS board around and got to the point (somehow) with < 60 nS values on all SQUETS.
    • Plugged all SQUETS and adjusted the boards to all phonon channels read values correctly (no loose pins). Didn't close IVC yesterday to check if the short returns.
    • It did return and resistance to ground is ~13 Mohm, very slowly going up. The time constant behavior still exist.
  • Priority Updates
  • General Updates
    • Himangshu
      • Elliott and I met to discuss the DMC simulation work today. I have received Macro from David, we checked basic simulation code steps on Elliott's (Narval- Canada server) and my SLAC opendemand cdms server. We go through the macro and identified steps that we don't understand fully. Plan is to read on those simulations steps, run a few events to reproduce the first plot in the slide and Find more details on T3 detector to understand how the code will be translated to be used for our study.
    • Shubham
    • Zach
    • Elliott
    • Scott