UMN CDMS Wiki archive
start › cdms › k100 › group_meetings › 20250918
Teleconference link: https://umn.zoom.us/j/9024497080?omn=95322888432
Meeting ID: 902 449 7080
Domestic: 1-408-638-0968 (US toll)
Access code:
  • Lab Update
    • Removed LED1 pin out of dib1, but led2 short to chassis still exist.
    • Visual inspection of led2 doesnt reveal anything noteable.
    • Cannot replicate the other phonon lines short to chassis at the bench.
    • Installed tower in the fridge again to see if we still see the short to chassis on other phonon lines.
    • Short to chassis still exist when checked at the breakout box. But I cannot see that while looking at the tower pins directly.
    • Suspecting stripline or SQUET issue, switched to another stripline and it vanished.
    • Reconnected to old stripline and it went away as well. Not sure what changed.
    • We will continue installing the stages.
    • Pump room power failure:
      • The power supply box connecting to He3 and He4 pump is not responding.
      • Bill visited briefly to discuss the issue and we both think it is some fuse blown out due to other electrical issues.
      • I got the diagrams ready and we will look into it today.
    • Radiation Safety Audit next week.
  • Priority Updates
  • General Updates
    • Himangshu
    • Shubham
    • Zach
    • Elliott
    • Scott