  • Lab Update - Run 82 on-going
    • Leak detector on IVC shows high leak!
    • Pumped out the IVC with turbo and leak detector in parallel to monitor leak level decreasing over time.
    • Closed IVC fridge valves on top and fridge and left overnight to see if levels again increase today.
    • If leak level remain same order of magnitude, we will start 1K again to see if that helped.
    • MC temp increase upto 7 K when turbo is pumping. After turbo turned off, it remains at 7K. Still and 1K warmer.