  • Lab Update - Run 82 on-going
    • Stopped 1K and inspected the cryopump. Refilled oil level and cleaned all seals.
    • Starting back 1K pot shows temperatures reaching 1.52 K which might indicate the pump is not effectively pumping before.
    • Yesterday restarted 1K pot, after LHe transfer, but again it shows 1.9 K temperature.
    • Pulled mixture back into the dump and started 1K shows 1.52 K.
    • But adding mixture push it quickly to >2 K and currently at 2.45 K, Still at 3.5K, MC 4.13 K.