Fridge Status : Fridge is at base temp of 45 mK,
On Friday, Paul suggested to repeat cycles of one-shot and adding mixture from dump to go lower in temperature.
We are at a stable base temperature of 45 mK. (earlier we get to 30-35 mK)
Tower is warm at 112 mK. Turned on DCRCs to tune SQUIDs.
Tower at 138 mK.
-
Electrical checks on channels show no shorts, but Side2pD feedback is not stable reading, might be a pin issue on the warmer side. Needs more checks. All other channels looks fine. Motor Coil resistances at ~ 36 ohm. (room temperature R ~ 150 ohm).
Running IbIs on phonon channels. But donot see any pulses yet.
Turned on new DAQ but L2Trigger showing error to start. Yan came in the late afternoon and found one L2Trigger old instance. Hopefully we can run it today.