UMN CDMS Wiki archive
start › cdms › k100 › group_meetings › 20240910
Teleconference link: https://umn.zoom.us/j/9024497080?omn=98454621599
Meeting ID: 902 449 7080
Domestic: 1-408-638-0968 (US toll)
Access code:
  • Lab Update - Run 81 paused
  • Priority Updates
  • General Updates
    • Himangshu
    • Shubham
    • Zach
    • Elliott
    • Jacob
  • Fun Update