  • Lab Update - Run 81 started
    • LHe transfer on Tuesday
    • We are at base temperature ~ 35 mK.
    • Tower is still cooling down.
    • Opening V2 valve for mixture circulation can reach 1K MC temperature.
    • VT(earlier V6) valve opened completely by turning the knob anti-clockwise to reach base. Needs stable hook for this valve.
  • Priority Updates
  • General Updates
    • Himangshu
    • Shubham
      • First dump of R79 series shows event repetition after 4 events. (A B C D A B C D ….)
      • This is strongly correlated with the fact that each MIDAS event in the first dump contains 4 triggers as opposed to later dumps where each MIDAS event contains about 126-128 triggers so in those dumps CDMS-event repetition occurs after 126-128 events.
      • For this run: we should aim for a fast turnaround analysis to check whether this issue persists.