  • Lab Update - Run 81 preparation
    • LN2 transfer on Friday. Mini-leak test suggested by Prof. Paul showed no leaks.
    • Final leak test done yesterday since bg levels were at 10^-6 mbar L/s (high).
    • Turbo pumping over the weekend helped to lower down the level.
    • During transfer stick removal, there was a leak signal but it seems to be not related to the indium seals.
    • LHe transfer completed yesterday.
    • Plan is to do another LHe transfer today and start circulation.
  • Priority Updates
    • People using jupyter notebook server on CDMS2: Don't forget to terminate the server after use. It does not shut down on its own and eats up the resources (Bad for other users).
    • Opportunity for detector operation training: see Ray's email.
  • General Updates
    • Himangshu
    • Shubham
      • Event duplication issue:
        • Elliot had noticed that R79 RQ data (and raw data) has duplicate events.
        • Investigation into this issue shows same ~127-128 events repeat for multiple times. The issue is in the raw data itself.
        • R76 looks fine.
        • We suspect same physics events have different trigger information therefore being treated as different CDMS events.
        • Different trigger settings for R79? Checking ODB keys.
        • Trying process the raw data with latest version of cdms software. It does not work.
    • Zach
    • Elliott
    • Jacob