  • Lab Update - Run 81 preparation
    • LN2 blowout on Thursday revealed leak towards the end of the He gas fill.
    • This is the bottom IVC seal leak and it is clearly seen on the leak detector.
    • We decided to warmup to replace this seal.
    • NaI array disassembled over the weekend to get ready to open the fridge.
    • Main Bath lowered yesterday and tent is erected.
    • Plan is to replace the indium seal, and pump back down today after leak testing.