  • Lab Update - Run 81 preparation
    • Proceeded with two more cryocleaning cycles to reach below 40 mbar of contaminents.
    • Rechecking electrical connections show ~ 50 nS of shorts in almost all channels. We prefer to keep it below 10 nS. Will check again at 77K if this changes.
    • Zach helped doing first mixture throughput test with the new procedure steps. ~ 3.8 mbar/min (roughly).
    • With a least count of 50 mbar on G1, it is hard to check changes with this rate. But we expect to see it better at 77K.