  • Lab Update - Run 80 ended
    • Found leak in the dump line. Traced it into the swagelok that was used to connect to the new dump line.
    • Made a new swagelok tube from scratches, thanks to Mike.
    • Leak testing the GHS system shows background level at 10^-7 mbar L/s.
    • Paul suggested to close cold trap ports and try background upto 10^-10 level.
    • Connections for He3 return line swap is ready.
    • Changes made today:
    • We will repeat leak testing and swap connections today. Prof. Paul has agreed to be there after 12 today.