  • Lab Update - Run 80 ended
    • IGH issue: might have been fixed
    • Nick Dent responded that the error shows issue with electrical circuits in IDR or IGH unit and they donot provide further support for repairs.
    • Tried unplugging the IDR from Goldman's lab (to plug it into ours) Proceeded with opening up IGH as well. Plugged it into ours.
    • IDR displaced SOS errors for a long time (~ 3 hrs) but then started settling down. Current state: Valves can be opened/closed. He4 pump can be started.
    • There is a clicking sound when we started the He4 pump, investigating if that can cause the IDR to fail again.
    • Plan is to recover the mixture and cooldown once the electronic locks are back online in the buidling.