  • Lab Update - Run 80 preparation
    • Connected Side1HV box and the two DCRCs using configuration as used in Run 76
    • DCRC3 squid tuning done. ChC donot show test signal. DCRC1 not showing test signal on all channels.
    • Swapped the two DCRCs to use DCRC3 for the other set of channels. Temp went up.
    • Removed DCRC1 from the setup. Again DCRC3 pC is not showing test signal.
    • DCRC3pB swapped going out of range when test signal is applied.

[image not archived: dcrc3pb_swapped.png]