  • Lab Update - Run 80 preparation
    • We are still at 77K! Encountered a few issues listed below.
    • Leak detector showing leak close to the Main Bath seal. Tried tightening screws more, but not helping much. Can still continue cooldown.
    • After LN2 blowout, P2 (Main Bath pressure) not going below 30 mbar (expected to go < 0.1 mbar). Is there LN2 remained inside but not leaving the system somehow like iced up conical fill line? Or is this issue connected with Issue 1?
    • Started He4 pump to switch with the scroll pump and saw smoke in the exhaust!
    • Asking Mike to see if he have any spare pumps to use for the time.
    • LHe Liquifier reached Linde's repair center yesterday. Will need 2 weeks to reach back to UMN.