SQUID tuning done for all phonon channels using DCRC3.
Cable 1 and Cable 3 shows good phonon pulses.
Cable 2 is responding to QET bias changes but no pulses seen yet. More noisy than other cables. Side2pE on this Cable.
Rough IbIs scans done by Guilia on Tuesday. Finer IbIs data taken on Wednesday.
-
Fridge Tower temp goes from 62 mK to 71 mK after connecting Cable 2, not on other cables. MC base is still at 30 mK. Need to figure out the cause of heating.
-
Switching back to two people LHe transfers with the Linde dewars