- Lab Update - Run 78
- After opening up the IVC bottom cap, there are two shorts (not on the phonon lines though). The Y-motor arm 1st pair and the endstops.
- After I started removing the SQUETS to look for shorts at the tower, only the Y-motor arm short remained.
- After I lower the tower, none of the shorts show up.
- Source mover didnt move at all! Source is still at the center.
- Phonon lines not getting biased is still not solved.
- Possible changes: Swap ZC-13 with ZC-31(earlier tested in Run 77)