- Lab Update - Run 78
- new DAQ Update:
- Power cycling the RevE board and re-establish phonon pointers helped to keep DAQ to continue without failure.
- Solder all IO cable slots on the adapter board and tried running all cables. Cannot use Cable2 with diverter board due to Male to Male DSUB connection issue, ordered Female Gender changer.
- 3 out of 5 channels showing traces on new DAQ while connecting Cable 1 and Cable 3. Turning on SQUID tuning Tools stops the run. Changing QET biases don't change baseline.
- old DAQ Update:
- Connected HV board to use only DCRC1 and DCRC3. Still can't change QET bias on few channels.
- Before warmup, we have two days. Plan will be to check HV bias on the detector and finding out max voltage. Source mover test to find out if the endstop strategy can be used to identify current position of the source, while its cold.
- Priority Updates
- General Updates
- Himangshu
- COMSOL Update: Simple Al mask pattern with housing+post results Slides
- Monitoring Page Update: Setup complete for Environmental Monitoring Status Page, also being used for the current CUTE run.
- Next task will be to add live alarm status on the main page to help experts check the alarm status quickly.